Skip to main content

Table 1 Limit of detection and enhancement factor of magnetic bead-enhanced RDTs relative to an unenhanced RDT

From: Direct transfer of HRPII-magnetic bead complexes to malaria rapid diagnostic tests significantly improves test sensitivity

Sample volume (μL)

LOD

(parsaites/μL)

Enhancement factor

5a

15.4

(Unenhanced)

25

7.9

1.9

50

5.5

2.8

100

1.7

9.1

250

0.73

21

  1. aControl RDT performed according to manufacturer’s protocol