Skip to main content

Table 2 Precision performance of the immunofluorescence test kit

From: Development and performance evaluation of a novel immunofluorescence chromatographic assay for histidine-rich protein 2 of Plasmodium falciparum

Time

Lot 1

Lot 2

Lot with lot CV (%)

Result

S/CO

Result

S/CO

1

Positive

1.75

Positive

1.81

6.04

2

Positive

1.69

Positive

1.89

3

Positive

1.60

Positive

1.98

4

Positive

1.70

Positive

1.91

5

Positive

1.73

Positive

1.88

6

Positive

1.69

Positive

1.95

7

Positive

1.78

Positive

1.85

8

Positive

1.66

Positive

1.87

9

Positive

1.77

Positive

1.71

10

Positive

1.64

Positive

1.82

Within lot CV (%)

3.39

4.10

  1. S/CO indicates the ratio of detected value to the cut-off value. S/CO < 1 is negative, S/CO > 1 is positive. The greater the S/CO values, the higher the HRP2 concentration