Table 1 Characteristics of Alethia® Illumigene Malaria kit and HumaTurb Loopamp™ Malaria PDT kit

From: Head-to-head comparison of two loop-mediated isothermal amplification (LAMP) kits for diagnosis of malaria in a non-endemic setting

 

Alethia® illumigene Malaria

HumaTurb Loopamp™ Malaria PDT

Mean time to result as measured in the study

44 min

60 min

Time to result as reported by the manufacturera

 < 45 min

50–55 min

Mean active time as measured in the study

4 min

10 min

Analysisa

Qualitative

Qualitative + turbidity graph and time to detection

Maximum samples per runa

10

16 (up to 94 if expanded with 6 HumaTurb A units)

Species identificationa

Not possible

Not possible

Validated for species detection

P. falciparum, P. vivax, P. ovale, P. malariae. Performance for P. knowlesi established using purified genomic DNA only; whole organism testing has not been performed

P. falciparum, P. vivax, P. ovale, P. malariae.

P. knowlesi validated in one study [20]b

Read out of the resultsa

Turbidity in Illumipro-10™ incubator or by eye

Turbidity in turbidimeter or fluorescence detected by eye

Limit of detectiona

2.0 p/µL for P. falciparum and 0.1 p/µL for P. vivax

1 p/µL

  1. aAs described by the manufacturer [14, 30, 32]
  2. b [20]